beautypg.com

American Diagnostic Corporation (ADC) ADview Modular Diagnostic Station User Manual

Page 43

background image

33

To replace the battery:

1. Remove the four screws securing the battery bay door.
2. Carefully remove the battery from the battery bay, being careful not to pull on the wires

attached to the battery terminals.

3. Disconnect the wires from the battery terminals.
4. The rechargeable battery contains lead. Please dispose of the old battery properly.
5. Connect the wires to the terminals of the replacement battery, ensuring the red wire is

attached to the red terminal and the black wire to the black terminal. If the wires are reversed,
no damage will occur, however the ADview will not operate. Be sure to use ADC item
number 9000BAT for the replacement battery in order to maintain optimum performance.

6. Re-secure the battery bay door with the four screws removed in step 1.
7. Connect the power supply to turn the device on and charge the replacement battery fully

before using.

If the rechargeable battery is disconnected for any reason, the device must be
connected to AC mains power via the power supply before the unit will turn on; this is
required even if the battery has been properly reconnected.

Disposal

This symbol indicates the device contains materials (such as electrical components) which are
hazardous. Please return to ADC for disposal.