Cannon Instrument miniPV-X User Manual
Page 76

CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
70
displayed on the VI analysis. The VI Matching function is always indi-
cated on the VISCPRO
®
status bar, whether or not a VI domain (viscom-
eter tube set) has been defined.
Configuration procedure
To configure the VI Matcher:
1. Load the VISCPRO
®
software.
2. Log on to the system as a Manager (a password may be required) to
access the VI Matcher function, which is security-protected.
NOTE
For more information on the security log-in procedure, see Security
Options, this chapter.
3. Select
VI Matching—General Configuration
from the
Configura-
tion
menu options. The
General Configuration
window will appear.
The VI Matcher General Configuration window
4. Click on the
Allow Matching of Invalid Samples
box if you wish to
permit VI matching of samples with flow times exceeding the
specified tolerance for repeatability.
NOTE
The VI Matcher does not match invalid samples UNLESS the Allow
Matching ... box is checked. The VI Matcher never matches unsuccessful
samples (samples with no flow time and a displayed kinematic viscosity of
“0”). If one of the samples ran successfully at one temperature but not at
the other, the good sample can still be used for VI matching if the failed
sample is rerun within the time frame selected for the VI Matching Date
Filter for the analysis.
5. Click on the
ADD
button to add a new VI configuration.