3. To calibrate temperature, enter the current temperature reading to the

nearest 0.01°C into the

Temperature:

field, subject to the following:

A decimal point MAY be included as an entry. It should be preceded
by at least two numbers or a "0".

The entered actual temperature must not cause a correction in
temperature greater than +/- 5.0 degrees.

Multiple or successive entries of correction for any temperature are
permissible. A newly entered corrected temperature is mathematically
compensated with the previously entered correction and is stored in
memory by the onboard microprocessor as the latest correction.

Data entry errors

The miniPV Series will not accept temperature reference data outside the
parameters noted previously. If the amount of the correction (delta to
displayed temperature) is too large, the correction will not be accepted.

General/Current Offsets

The

General Bath Offset

is factory-preset value affecting temperature

control at all temperatures and should not ordinarily be changed. The

Current Offset

is the temperature adjustment applied to a particular

temperature to conform bath temperature control to your temperature
reference . Always click the radio button for the

Apply As Current

Option

when calibrating for a specific temperature.

Training sensors

It is necessary to train instrument sensors in preparation for sample
testing at a specific temperature. The training procedure establishes the
proper power levels for sensor detection of the sample meniscus. The
sensor training data is then stored by the miniPV Series instrument, and
recalled when testing at that temperature. The sensor training procedure
requires approximately 30 minutes (longer at lower temperatures).

NOTE

You will be prompted to train sensors by the instrument hardware or
software whenever a new temperature is selected for miniPV Series
operation. You will be unable to test samples until the training has been