Training sensors – Cannon Instrument miniPV-X User Manual
Page 36

CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
30
3. To calibrate temperature, enter the current temperature reading to the
nearest 0.01°C into the
Temperature:
field, subject to the following:
A decimal point MAY be included as an entry. It should be preceded
by at least two numbers or a "0".
The entered actual temperature must not cause a correction in
temperature greater than +/- 5.0 degrees.
Multiple or successive entries of correction for any temperature are
permissible. A newly entered corrected temperature is mathematically
compensated with the previously entered correction and is stored in
memory by the onboard microprocessor as the latest correction.
Data entry errors
The miniPV Series will not accept temperature reference data outside the
parameters noted previously. If the amount of the correction (delta to
displayed temperature) is too large, the correction will not be accepted.
General/Current Offsets
The
General Bath Offset
is factory-preset value affecting temperature
control at all temperatures and should not ordinarily be changed. The
Current Offset
is the temperature adjustment applied to a particular
temperature to conform bath temperature control to your temperature
reference . Always click the radio button for the
Apply As Current
Option
when calibrating for a specific temperature.
Training sensors
It is necessary to train instrument sensors in preparation for sample
testing at a specific temperature. The training procedure establishes the
proper power levels for sensor detection of the sample meniscus. The
sensor training data is then stored by the miniPV Series instrument, and
recalled when testing at that temperature. The sensor training procedure
requires approximately 30 minutes (longer at lower temperatures).
NOTE
You will be prompted to train sensors by the instrument hardware or
software whenever a new temperature is selected for miniPV Series
operation. You will be unable to test samples until the training has been