4. Attach a shorter RJ-45 cable to one of the twin RS-485 connectors

on the ear panel of the first miniAV bath. Then attach the other end
of the cable to one of the twin RS-485 connectors on the ear panel of
the adjacent Bath Unit.

5. Continue the procedure from the previous step for each bath con-

secutively until you have completed RS-485 connections between
each Bath Unit and the previous unit in the sequence (see diagram).

------

------

------

6. For each networked instrument, extend the left ear panel and locate

the Instrument Address access on the interior side of the panel. The
adjust pot is inside the aperture. Select the desired instrument address
(1-9) using a small flat-bladed screwdriver. Make certain to select a
DIFFERENT address for each miniAV instrument.

7. Restore power to each miniAV.