beautypg.com

Cannon Instrument miniPV-X User Manual

Page 65

background image

59

CANNON

®

miniPV-Series Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.a— January, 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

You may evaluate the accuracy of the new calibration constants by
comparing the known viscosity for the standard with the calculated
viscosity for each standard selected for the new calibration. This differ-
ence is displayed as a percentage in the last column (% Diff) of the
updated sample list box.

Show Calibration Data

The

Show Calibration Data

button is a toggle that will permit you to

compare current and new calibration data and constant information.
Press the

Show Calibration Data

button once to view the current

calibration information. Note that the button name will change to

Display

Sample Data

. Press the button again to display the new calibration data.

When/if you are satisfied with calculation results, click on

Update Tube

to save calibration values for the current configuration. The new data will
become the current data.

If the difference is still too great, repeat the calibration process until the
difference falls within acceptable limits (refer to the precision statement
in ASTM D445).

The Calibration window

NOTE

You must click Update tube before changing bulbs if you wish to save the
new calibration constants for that bulb. If you close the
Calibration window
without updating the tube, the new calibration information will not be
saved.