beautypg.com

Service menu – Cannon Instrument miniPV-X User Manual

Page 66

background image

CANNON

®

miniPV-Series Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.a— January, 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

60

Service menu

To access the service menu items, click

Service

from the VISCPRO

®

primary display and select the desired instrument.

The

Monitor

window permits the user to view current operational values

for the miniPV Series instrument. These settings may be helpful to

CANNON

®

technical personnel when troubleshooting difficulties with

the miniPV Series.

Select

Wash viscometer tube

to open the

Wash viscometer tube

window.

Click

Begin Wash

to initiate the wash cycle. The

Status:

indicator will

display and update current actions.

Bypass .... option

To begin washing an evacuated sample cup, click the option box to make
certain that the

Bypass initial cup evacuation

option is selected.

CAUTION

Do not use this option with a filled cup or the cup will overflow.

When the

Wash

cycle has been completed, click

Done

to exit the

Wash

viscometer tube

window.

NOTES

The Wash operation cannot be executed if the tube is currently running a
sample. To terminate the Wash cycle at any time, click
Abort Wash.

CAUTION

The Wash viscometer tube function overrides automatic miniPV Series
software operations. Verify that the tubes are clean and dry before initiat-
ing computer-controlled sample testing.