Service menu – Cannon Instrument miniPV-X User Manual
Page 66
CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
60
Service menu
To access the service menu items, click
Service
from the VISCPRO
®
primary display and select the desired instrument.
The
Monitor
window permits the user to view current operational values
for the miniPV Series instrument. These settings may be helpful to
CANNON
®
technical personnel when troubleshooting difficulties with
the miniPV Series.
Select
Wash viscometer tube
to open the
Wash viscometer tube
window.
Click
Begin Wash
to initiate the wash cycle. The
Status:
indicator will
display and update current actions.
Bypass .... option
To begin washing an evacuated sample cup, click the option box to make
certain that the
Bypass initial cup evacuation
option is selected.
CAUTION
Do not use this option with a filled cup or the cup will overflow.
When the
Wash
cycle has been completed, click
Done
to exit the
Wash
viscometer tube
window.
NOTES
The Wash operation cannot be executed if the tube is currently running a
sample. To terminate the Wash cycle at any time, click Abort Wash.
CAUTION
The Wash viscometer tube function overrides automatic miniPV Series
software operations. Verify that the tubes are clean and dry before initiat-
ing computer-controlled sample testing.