these samples should also be covered before analysis. Most lubricating
oils, because they are manufactured at high temperatures and low pres-
sures, do not contain significant volatile components and can be analyzed
without being covered.
The PolyVISC SPS Solution Preparation System from Cannon Instrument
Company is a semi-automated solution preparation system that can be
used with the minPV instruments. Contact Cannon for more information.
NOTE
For additional information on software data entry features, including
multiple sample selection and cut & paste options, see Testing samples—
software options in Chapter 4.
Running check standards
If
Verify Known KV
is selected as the test option for a sample, the
Viscosity Action
window will open automatically. Enter the neces-
sary check standard data, including the
Check Standard viscosity
from the standard bottle, and click
OK
to complete data entry. To
revise or confirm standard data, right-click on the desired sample ID
from the list box and select
Configure
from the popup menu choices.
NOTE
See Chapter 4 for additional information on the Viscosity Action window.
9. Continue entering sample information for all desired trays. When
sample ID data entry is complete, check the
Tray Status
window to
verify that all trays are ready for testing.
10. Click on the
RUN
button at the bottom of the
Sample Input wind
ow.
The
Select Trays
window will open if more than one miniPV Series
instrument is online.
11. If necessary, click on the check box(es) to select the desired pre-
pared “tray” (sample sequence) for automatic testing. For the
miniPV Series, a tray corresponds to the miniPV Series instrument.
Then click
OK
to begin the miniPV Series test(s).
Pausing a test
To temporarily halt testing for a given tube/sample, click the
Pause
button from the
Sample Input wi
ndow. Then select the desired tray
(miniPV Series instrument) and pause action(s) from the
Select Trays
window (
Pause Now
will immediately pause test actions;
Pause after
current sample
will pause testing after the current test is complete).
Click
OK
to pause testing for the selected tray(s).
NOTE
If the test was paused prior to the initiation of the Wash cycle, drop time
data for that sample will be discarded.
Resuming a test
To resume test actions for paused trays, click the
Resume
button from
the
Sample Input wi
ndow. Then select the desired trays from the
Select
Trays
window. Click
OK
to resume sample testing (see note above).