3. Select the desired unit of temperature measurement (°C or °F).
4. Type the desired temperature in the
Bath Temp:
field. Acceptable
values are any numbers between 20°C and 100°C. If necessary, you
may use the decimal point key to input temperature to the nearest
0.01°C. Press
OK
to save the temperature setting.
NOTES
The miniPV Series will not allow sample testing until the instrument has
equilibrated temperature at the test temperature.
To cancel temperature selection before completing the procedure, click
Cancel.
After the target temperature has been set, the miniPV Series bath
temperature will be adjusted to the target temperature and the bath will
equilibrate at the test temperature. The VISCPRO II software always
displays the current and target bath temperature in the
Sample Input
wi
ndow.
Checking wash functions
To ensure function of the solvent wash system, follow the procedure
below:
1. Place an empty vial in the sample tray position 1.
2. Select
Wash viscometer tube
to open the
Wash viscometer tube
window.