Viewing/editing setup information – Cannon Instrument miniPV-X User Manual
Page 17

11
CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Viewing/editing setup information
NOTE
For first-time installation, make certain that the factory-prepared Configura-
tion file has been copied to the Default Data directory per instructions on
page 7 of manual. If your instrument has already been configured, you can
use the instructions in this section of the manual to check or, if necessary,
change the instrument settings.
1. Click
Configure
from the VISCPRO
®
menu bar.
2. Select your instrument group and instrument from the list of available
instruments.
3. Select
Tray Settings: Tube and Bath
from the list of configuration
options.
The Tray Settings: Tube and Bath
window will appear.
The Tray Settings: Tube and Bath window
The
Tray Settings: Tube and Bath
window contains setup information
for each tube associated with your instrument. On multiple-tube CAN-
NON instruments, you can click on the tube tabs to see the setup infor-
mation for different tubes.
4. Verify that the tube serial number (
Tube S/N
) is correct. If it is not,
input the correct serial number in the text box. Each tube should have
a unique serial number!
5. Verify the presence of calibration
values (
C
and
E
) for each bulb. If
calibration data is not available, the
default values are C=1 and E=0.
Sample calibration constants