Minipv series calibration equations – Cannon Instrument miniPV-X User Manual
Page 41

35
CANNON
®
miniPV-Series Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.a— January, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
in effect until the instrument is recalibrated at that same temperature, or
until instrument and tray settings from a previously-saved instrument
configuration are restored.
To avoid losing valid calibration data,
CANNON
®
recommends saving
the instrument configuration settings after completing a new calibration
(see Saving a configuration, chapter 4). The name for the instrument
configuration might ordinarily identify the instrument, the calibration date,
and the temperature:
EXAMPLE:
miniPV JAN10 40C
After the instrument configuration has been saved, calibration data may
be restored at any time using the
Restore Instrument & Tray Settings
option from the Configure menu. Note that the
Restore Instrument
Settings
and
Restore Tray Settings
check boxes must both be selected
to restore instrument calibration data:
NOTE
The Restore Instrument Settings and Restore Tray Settings options are
security-protected at the Manager level.
miniPV Series calibration equations
Viscosity calculation equation In the miniPV Series, viscosity is calculated by using the viscosity calcu-
lation equations as described in ASTM Methods D445 and D 446.
Because samples running in the miniPV Series can have short flow times,
the kinetic energy correction term (E / t
2
) is used as prescribed in the
aforementioned ASTM methods. The VISCPRO
®
software for Win-
dows
®
performs all necessary calculations to derive constants
C
and
E
.
These constants are used by the software when determining values for
kinematic viscosity using sample drop times measured by the miniPV
Series.
Kinematic Viscosity equation: v = C t - E / t
2
Where: v
=
kinematic viscosity (cSt)
C
=
calibration constant (cSt/s)
E
=
kinetic energy correction (cSt·s
2
)