beautypg.com

Appendix a–multiple unit configuration, Introduction, Kit components – Cannon Instrument miniPV-X User Manual

Page 127: Procedure

background image

121

CANNON

®

miniAV

®

-HX Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0— January, 2011;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CHAPTER

A

APPENDIX A–MULTIPLE UNIT

CONFIGURATION

Introduction

The miniAV instrument is capable of being networked with the controlling
computer via available RS-485 connections. As many as four miniAV
instruments may be connected to a single computer. Follow the instruc-
tions in this appendix to complete assembly of the multi-unit networking
kit, available from

CANNON

®

.

Kit components

The multi-unit interface kit contains an RS-232-to-RS-485 converter and
RJ-45 cables capable of networking up to four CAV Bath Units to a
single computer.

Procedure

Using the interface kit components, complete the multi-unit configuration
as follows:

Connecting components

NOTE

The interface kit was designed for Bath Units placed in an adjacent
configuration. Custom configuration with longer RJ-45 cables is possible.

Make certain to install the VISCPRO

®

software on the controlling

computer. Complete all component assembly before proceeding with RS-
485 networking.

Completing RS-485 connections

CAUTION

1. Make sure the power is removed from the computer and all instru-

ments to be configured.

2. Locate the RS-485 converter and required RJ-45 connecting

cable(s).

3. Attach the RS-485 converter directly to the computer via the stan-

dard 9-pin RS-232 connector. Then attach the longer RJ-45 cable
from the converter to the RS-232 connector on the serial communica-
tions interface on the ear panel of the first miniAV instrument.