Cancel to exit the Save Calibration Data window without saving the
information. The calibration certificate will be printed after you exit the
Save Calibration Data window.
CANNON Instrument Company recommends saving instrument and
tray settings each time a new calibration is performed.

Select

Save Instrument and Tray Settings

to save certain current

instrument information, including instrument specifications and calibration
data (

Instrument Settings

), basic and advanced operational parameters

(

Tray Settings

), and previously-entered Sample ID information.

When you click the

Save Instrument and Tray Settings

menu item, the

Save Configuration

window will open.

The Save Configuration window

Choose a name for the Configuration by entering it in the

Save As:

field.

You may also click on existing (saved) Configuration by clicking on the
Configuration name in the

Existing Configurations

list box.

When you have selected a Configuration name, click OK to save the
Configuration and exit the Save Configuration window. Or click Cancel to
exit the window without saving the Configuration. You can also delete
existing configurations by selecting them from the Existing Configurations
list box and pressing the Delete button.

Select Restore Instrument and Tray Settings to restore certain

instrument information, including instrument specifications and calibration
data (Instrument settings) basic and advanced operational parameters
(Tray Settings), and previously-entered Sample ID information.

When you click the

Restore Instrument and Tray Settings

menu

option, the

Restore Configuration

window will open.