The Calibration window
Compare this value to your archive of the calibration constant data (if
available) for bulb 1. The values should be identical.
3. If they are not, place your cursor in the appropriate field, delete the
entry, then type the correct values for the constant in the text box.
4. Click
Update tube
.
5. Use the bulb spin controls
to select the other bulb(s) for which
you noted calibration constant errors. Input the correct values for
each. Make certain to click
Update Tube
after you have corrected
C
and
E
calibration values for each bulb BEFORE selecting the next
bulb.
6. When you have entered corrected constant values for each bulb on
both tubes, click
to exit the
Calibration
window.
You have verified the software configuration of VISCPRO
®
. To test
samples with your instrument, follow the instructions in Chapter 2. For
additional details regarding operating procedures for your instrument or
software, consult the appropriate section of this manual.
Checking Wash and Advanced Settings
To check Wash Settings for your miniPV Series instrument, click
Config-
ure
from the VISCPRO menu options and select the desired instrument.
Then click
Tray Settings: Wash
. The
Tray Settings: Wash
window
will appear.