Make certain to enter the correct check standard viscosity (in centistokes)
BEFORE running the standards. Incorrect data entry may skew future
calibration results.
5. Click
Run
, then select the appropriate sample trays to test the
standards.
6. When testing is completed, use the VISCPRO
®
software to log on as
a Manager (
Main/Log In
) in order to access the calibration menu.
7. Select the desired miniPV Series instrument from the
Configure
menu; then select
Calibration
.
The
Calibration
window will open.
8. Select the date range for acceptable calibration data by using the
No
older than [x]
days spin controls
to set the date search param-
eters (if “
1
” is selected, only data from the standards run in the last
24 hours will be listed). Any check standards from the time frame
you specified which are appropriate for calibration calculations will
appear in the window list box.
The miniPV Series Calibration window
9. Use the spin controls
to select the desired
Bulb
for calibration.