9.11 Updating Firmware
To update firmware of the
Counterpart
, you must have Revolution installed and a .hex file on your computer.
Visit
www.ricelake.com
to download this free configuration software and the latest .hex file.
If the hex file is the same version as is currently in the indicator, the firmware update will not reset the
configuration. This is helpful if the firmware becomes corrupt and you want to reload the same firmware.
However, it is always recommended that configuration be backed up in Revolution to avoid any data loss.
1. Unplug power to the
Counterpart
and remove top from the base of the unit.
2. Wire the serial cable to the
Counterpart
’s Port 1. Refer to Table 9-7 for cable connection settings.
3. Install jumpers on JP5 and JP6. See Figure 2-4 on page 9.
4. Plug in power to the
Counterpart
and press the indicator’s
Power
button.
5. With Revolution open, begin a new configuration file for the
Counterpart
.
6. Click
Update
Counterpart
Firmware
.
Figure 9-4. Revolution Screen.
7. The Rice Lake Updater screen appears. Specify the COM port the
Counterpart
is connected to, and click
the ellipses (...) to browse to and select the desired .hex file.
DB9 Cable
Connector (J5)
Pin/Color
Signal
In/Out
Description
Pin
Description
In/Out
1/Brown
DCD
In
Data Carrier Detect
NC
NA
NA
2/Red
RxD
In
Receive Data
3
Tx
Out
3/Orange
TxD
Out
Transmit Data
4
Rx
In
4/Yellow
DTR
Out
Data Terminal Ready
5
DTR
In
5/Green
GND
--------
Ground
1 or 2
Gnd
--------
6/Blue
DSR
In
Data Set Ready
NC
NA
NA
7/Purple
RTS
Out
Request to Send
6
RTS
In
8/Gray
CTS
In
Clear to Send
NC
NA
NA
9/Black
RI
In
Ring Indicator
NC
NA
NA
Table 9-7. Cable connection settings
Note