49
4.0
Calibration
The
Counterpart
can be calibrated using the front panel, EDP commands, or Revolution
®
.
Calibration can be performed in two places within the menu: the CALIBR menu shown in Figure 4-1 and the
SCALE submenu shown in Figure 3-6 on Page 22. The CALIBR menu shown in Figure 4-1 is a “quick
access” calibration; for more in-depth scale setup and calibration, use the menus found under
SETUP
»
CONFIG
»
SCALE (see Figure 3-6 on Page 22).
Calibration consists of the following steps:
•
Zero calibration
•
Entering the test weight value
•
Span calibration
•
Optional five-point linearization
•
Optional rezero calibration for test weights
using hooks or chains.
The
Counterpart
requires the WZERO and WSPAN points to be calibrated. The linearity points are
optional, but must NOT duplicate zero or span. During calibration, the TARE(Enter) key acts as a data entry
confirmation key. The soft calibrate key accepts that value.
Figure 4-1. Calibration (CALIBR) Menu
4.1
Front Panel Calibration
1. Press the
MENU SETUP
button, then press
u n t i l t h e d i s p l a y r e a d s
C A L I B R
( s e e
2. Press
TARE(Enter)
or to go to the scale to be
calibrated. Press
TARE(Enter)
or to go to
WZERO
. Press
TARE(Enter)
once again to
display a value and press the
Calibrate softkey
to
Calibrate
WZERO.
If you do not want to calibrate zero, press
Menu to exit.
3. The indicator displays
CALIBRATING
while
calibration is in progress. When complete,
press
TARE(Enter)
or to advanced to the next
prompt (
WVAL
).
To view the new A/D count, repeat Step 3,
but instead of pressing TARE(Enter) while
viewing the value, press MENU SETUP to
exit.
4. With
WVAL
displayed, press
TARE(Enter) or
to display or edit the stored calibration weight
value. Press
TARE(Enter)
to store the
WVAL
value and advance to
WSPAN
.
5. With
WSPAN
displayed, press
TARE(Enter)
or
to view the A/D count. Place test weights on
the scale equal to WVAL. Press the
Calibrate
softkey
to calibrate WSPAN.
If you do not want to calibrate span, press
Menu to exit.
6. After
Calibrate softkey
is pressed, the indicator
displays
CALIBRATING
when complete. Press
TARE(Enter)
or and the
SaveExit softkey
or
proceed to
WLIN
.
To view the new A/D count, navigate back to
the WSPAN, repeat Step 6; however, instead
of pressing Calibrate softkey while viewing
the value, press Save/Exit to exit.
Note
Note
WZERO
Calibrating
Calibrating
Display and edi
Press the calibrate
softkey to calibrate
zero.
t
test weight value
WVAL
WSPAN
Press calibrate
softkey to remove
offset from zero from
zero and span
calibrations
REZERO
WLI N
POINT 1
POINT 2
POINT 3
POINT 4
POINT 5
Same as POINT 1
WGT 1
CAL1
Calibrating
Press the calibrate
softkey to calibrate
span.
XXXXXXX
ID
DISPLAY
TEST
SETUP
AUDIT
CALIBR
Press Menu
to cancel
Press Menu to
cancel
Previous A/D raw
counts are shown
Previous A/D raw
counts are shown
SCALE 1
SCALE 2
SETPTS
Note
Note
Note
Note