Appendix D – Configuring the Totalizer
Page 5 of 5
Use the up/down arrow keys to select
Low cutoff
and touch
ENTER
. You will be prompted with
cut
→
→
→
→
0.0
←
←
←
←
. This
feature allows you to enter a minimum flow rate at which the
totalizer should cut out and stop totalizing. This is for
transmitters that become very inaccurate at low flow rates.
Since ours is an open channel meter, we will not program a
low cutoff. Leave it set at
0.0
and touch
ENTER
.
21.
You should now be back in the menu beginning with
Point
tag
. Use the up/down arrow keys to select
Chart scale
and
touch
ENTER
. You will be prompted with
chart scl= A.
Use the up/down arrow keys to select
A
and touch
ENTER
.
Note
: Since the range of the flow meter is 0-50 MGD and
we will only be totalizing for a maximum of one day, the
totalizer will not exceed 50 so we can use the same scale.
If we were totalizing over long periods of time, we would
have to put in an appropriate scale for the totalizer channel.
Touch
EXIT
. You will be prompted
keep setup?
. Touch
YES
and you will be returned to the main program menu.
Touch
EXIT
several times until you are completely out of all
menus.
You have now successfully configured a totalization
channel! Use the
VIEW
button to select a particular display
to verify your work.
23.
22.
24.
You will now be back at the Reset Control menu. Use the
up/down arrow keys to select
Reset print
and touch
ENTER
. You will be prompted with
print=NO*
. Touch the
YES
key then touch
ENTER
. This enables the Recorder to
post the totalizer reset information to the alarm/event page
which is where you would view your daily totals.
Touch
EXIT
to get back to the menu beginning with
Point
tag
.