Monarch Instrument DC 2000 User Manual
Page 132
![background image](/manuals/598554/132/background.png)
Appendix A – Configuring a 1-5Vdc input for a 0-250 psig pressure transducer
Page 4 of 6
Using the up/down arrow keys select
Chart scale
and touch
ENTER
. A menu prompting you to choose a chart scale will
appear (
chart scl=
). Using the up/down arrow keys you can
select letters
A-H
for a total of 8 different scales. For this
example select chart scale
A
and touch
ENTER
. Then
touch
EXIT
.
You will now be prompted with
keep setup?
Touch
YES
and you will be returned to the main program window.
The Chart/Pens menu will be displayed. Using the up/down
arrow keys select
Scales
and touch
ENTER
.
Use the up/down arrow keys to select
Chart/Pens
and touch
ENTER
.
16.
17.
18.
19.
Using the up/down arrow keys select
Eng units
and touch
ENTER
. The alpha-numeric keypad will now be displayed
allowing you to enter up to five characters to describe the
engineering units for this particular point. There are
multiple pages of text and symbols that can be used. Use
the
PAGE
,
PAGEBK
and
SHIFT
keys to access different
characters. For this example (0-250 psig) enter
PSIG
and
touch
ENTER
.
15.