LaMotte SMART 2 User Manual
Page 95

PROCEDURE–FREE CHLORINE
1.
Press and hold
ON
button until colorimeter turns on.
2.
Press
ENTER
to start.
3.
Press
ENTER
to select TESTING MENU.
4.
Select ALL TESTS (or another sequence containing 15 Chlorine)
from TESTING MENU.
5.
Scroll to and select 15 Chlorine from menu.
6.
Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
7.
Insert tube into chamber, close lid and select SCAN BLANK
.
8.
Remove tube from colorimeter and pour off all but a sufficient amount of
sample water to cover a tablet. Add one *Chlorine DPD #1 Instrument
Grade Tablet (6903A). Crush tablet with a tablet crusher (0175), then
add sample water until tube is filled to 10 mL line. Cap tube and shake
until tablet has dissolved. Solution will turn pink if free chlorine is
present. Wait 15 seconds, but no longer than 30 seconds. Mix.
9.
Insert tube into chamber, close lid and select SCAN SAMPLE.
PROCEDURE–COMBINED CHLORINE
10.
Add one *Chlorine DPD #3 Instrument Grade Tablet (6197A) to sample
from Step 8 above. Crush tablet with tablet crusher (0175). Cap tube and
shake until tablet dissolves. An increase in color represents combined
chlorine.
þ NOTE: For wastewater samples, Standard Methods for the Examination of
Water and Wastewater recommends waiting 2 minutes for full color
development.
11.
Insert sample into chamber, close lid and select SCAN SAMPLE. Record
result as Total Chlorine.
12.
Subtract free chlorine reading from total chlorine reading to obtain
concentration of combined chlorine.
13.
Press the
OFF button to turn off the colorimeter or press the EXIT button
to exit to a previous menu or make another menu selection.
Smart2 TEST PROCEDURES 4.08
Chlorine–Bromine–Iodine 3/8