Blank done, 09 ppm – LaMotte SMART 2 User Manual

Page 30

background image

Press the

Q/ENTER button to select

002 Aluminum

.

ALL TESTS

* 002 Aluminum

003 Ammonia - NLF

004 Ammonia - NLS

The SMART2 Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press the

Q/ENTER

button to scan blank.
NOTE: Do not keep the button depressed.

002 Aluminum

* Scan Blank

The screen will display Blank Done for about
1 second. Scan Sample will be positioned next
to *.

002 Aluminum

Blank Done

* Scan Blank

Place the reacted sample in the chamber, close
the lid and press the

Q/ENTER button to scan

sample. The colorimeter will scan the sample and
the results screen will appear.

002 Aluminum

* Scan Sample

Record test result. To repeat the test, press the

Q/ENTER button to scan the sample again. The

last blank scanned is used to zero the colorimeter
for repeated scans. A different blank can be used
by pressing the

s button to scroll back to Scan

Blank

and then scanning another blank. Scroll

with the

t or s buttons and make another

selection with the

Q/ENTER button. The %T or

Absorbance of the last test can be viewed by
choosing %T/Abs. Press the

EXIT button to

escape to previous menus.
NOTE: The menus loop in this screen so either
the

s or t buttons will lead to the menu

selection needed.

002 Aluminum

0.09 ppm

* Scan Sample

Next Test

Previous Test

%/Abs

Calibrate

Scan Blank

30

SMART2 COLORIMETER 1.07