LaMotte SMART 2 User Manual
Page 150
PROCEDURE
Use 10 mm square cell adapter.
1.
Press and hold
ON button until colorimeter turns on.
2.
Press
ENTER to start.
3.
Press
ENTER to selct TESTING MENU.
4.
Select ALL TESTS (or another sequence containing
37 Cyanuric-UDV
) from TESTING MENU.
5.
Scroll to and select 37 Cyanuric-UDV from menu.
6.
Rinse a clean vial (0156) with sample water.
7.
Use the syringe (1184) to add 3 mL of sample to the vial.
8.
Insert the vial into chamber, close lid and select SCAN BLANK.
9.
Remove vial from colorimeter.
10.
Use the syringe (1184) to add 3 mL of sample to a Cyanuric Acid UDV
vial (4313).
11.
Invert the vial 3 times to mix.
12.
Wait 2 minutes.
13.
Invert vial 3 more times to mix.
þ NOTE: Firmly tap side of vial 5-10 times to remove all air bubbles.
14.
Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.
15.
Press
OFF button to turn the colorimeter off or press EXIT button to exit
to a previous menu or make another menu selection.
þ NOTE: For best possible results, a reagent blank should be determined to
account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents is obtained.
UDVs from opened pouches should be used promptly. Store unused vials from
opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the
reagent. Generally, UDVs stored in the bag should be used within 10 days if
the humidity is less than 50% and within 5 days if humidity is greater than
50%. The Foil Storage Bag contains a desiccant pack.
Cyanuric Acid–UDV 2/2
Smart2 TEST PROCEDURES 2.04