CALIBRATING LaMOTTE PRE-PROGRAMMED TESTS

The LaMotte Pre-Programmed Tests have been pre-calibrated. Recalibration of
the pre-programmed tests by the user is not possible. However, a procedure to
standardize the calibration can be performed to obtain the most accurate
readings or to meet regulatory requirements.

The LaMotte Pre-Programmed tests are standardized with one standard
solution. To standardize over the full range of the test, the concentration of the
standard should be chosen from the high end of the range. Alternatively, if
samples do not cover the full range of the test, a standard should be chosen
that is close to the concentration of the samples.

The standardization procedure should be followed as often as required by
regulations and laws for compliance monitoring.

In the example below the Aluminum calibration will be standardized.

Prepare a standard solution to be tested. Use 0.10 ppm aluminum.

Use the

s or t button to scroll to

002 Aluminum

. Follow instructions in the

SMART2 Manual for testing the aluminum
standard. Scan the blank.

002 Aluminum

* Scan Blank

The screen will display Blank Done for about
1 second. Scan Sample will be positioned next
to *.

002 Aluminum

Blank Done

* Scan Sample

Place the reacted sample in the chamber, close
the lid and press

Q/ENTER to scan sample. The

result will be displayed.

002 Aluminum

* Scan Sample

The displayed result can now be standardized.
Use the

s or t buttons to scroll to

Calibrate

. Press

Q/ENTER to select.

002 Aluminum

0.09 ppm

* Scan Sample

Next Test

Previous Test

%T/Abs

Calibrate

Scan Blank

SMART2 COLORIMETER 1.07