PROCEDURE

1.

Press and hold

ON

button until colorimeter turns on.

2.

Press

ENTER

to start.

3.

Press

ENTER

to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 78 Phosphate L)
from TESTING MENU.

5.

Scroll to and select 78 Phosphate L from menu.

6.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

7.

Insert tube into chamber, close lid and select SCAN BLANK.

8.

Remove tube from colorimeter. Use 1.0 mL pipet (0354) to add 1.0 mL of
*Phosphate Acid Reagent (V-6282). Cap and mix.

9.

Use the 0.1 g spoon (0699) to add one measure of *Phosphate Reducing
Reagent (V-6283). Cap and shake until powder dissolves. Wait 5 minutes
for full color development. Solution will turn blue if phosphates are
present.

10.

At end of 5 minute waiting period, mix, insert tube into chamber, close lid
and select SCAN SAMPLE. Record result.

11.

Press

OFF button to turn colorimeter off or press EXIT button to exit to a

previous menu or make another menu selection.

Smart2 TEST PROCEDURES 2.04

Phosphate–LR 3/3