Iron – LaMotte SMART 2 User Manual
Page 169
IRON
BIPYRIDYL METHOD • CODE 3648-SC
QUANTITY
CONTENTS
CODE
30 mL
*Iron Reagent #1
*V-4450-G
5 g
*Iron Reagent #2 Powder
*V-4451-C
1
Pipet, 0.5 mL, plastic
0353
1
Spoon, 0.1 g, plastic
0699
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
Most natural waters contain some iron. Its presence may vary from small traces
to very large amounts in water which is contaminated by acid mine wastes. For
domestic use, the concentration should not exceed 0.2 ppm and for some
industrial applications not even a trace of iron can be tolerated. There are
many means available for removing or reducing the iron content. Water
softening resins are effective for removing small amounts of iron and special
ion exchange materials are selective for iron removal. High concentrations of
iron can be removed by such chemical processes as oxidation and lime or
lime-soda softening. Because of the many means of removing or reducing the
amount of iron in water, the particular method employed will depend largely
on the form of iron which is present and the end use of the treated water.
APPLICATION:
Drinking, surface and saline waters; domestic and industrial
wastes.
RANGE:
0.00–6.00 Iron
METHOD:
Ferric iron is reduced to ferrous iron and subsequently forms
a colored complex with bipyridyl for a quantitative measure
of total iron.
SAMPLE
HANDLING &
PRESERVATION:
The sample container should be cleaned with acid and
rinsed with deionized water. Addition of acid to adjust the
sample to pH 2–3 will prevent deposition of iron on the
container walls. Samples should be analyzed as soon as
possible.
INTERFERENCES:
Strong oxidizing agents interfere, as well as copper and
cobalt in excess of 5.0 mg/L.
Smart2 TEST PROCEDURES 2.04
Iron–Bipyridyl 1/2