PROCEDURE

1.

Press and hold

ON

button until colorimeter turns on.

2.

Press

ENTER to start.

3.

Press

ENTER to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 77 Phenol)from
TESTING MENU

.

5.

Scroll to and select 77 Phenol from menu.

6.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

7.

Insert tube into chamber, close lid and select SCAN BLANK.

8.

Remove tube from colorimeter. Use the 0.1 g spoon (0699) to add one
measure of Aminoantipyrine Reagent (7825-C). Cap and mix.

9.

Use the plain pipet (0352) to add 4 drops of *Ammonium Hydroxide
Solution (7826). Cap and mix.

10.

Use the 1 mL pipet (0354) to add 2 mL of *Potassium Ferricyanide
Solution (7827). Cap and mix. Solution will almost immediately develop
a reddish hue if phenols are present.

11.

Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.

12.

Press

OFF button to turn colorimeter off or press EXIT button to exit to a

previous menu or make another menu selection.

Phenol 2/2

Smart2 TEST PROCEDURES 12.07