beautypg.com

Phosphorus, total–high range – LaMotte SMART 2 User Manual

Page 229

background image

PHOSPHORUS, TOTAL–HIGH RANGE

MOLYBDOVANADATE METHOD WITH ACID PERSULFATE
DIGESTION • CODE 4025

QUANTITY

CONTENTS

CODE

25

*Total Phosphorus Acid Reagent Tubes

*4035-G

60 mL

Deionized Water

5115PS-H

5 g

*Digestion Reagent Powder

*4036- C

2 X 30 mL

*Total Phosphorus HR Hydroxide Reagent

*4037-G

30 mL

*Total Phosphorus HR Indicator Reagent

*4039-G

1

Spoon, 0.15 g

0727

3

Pipets 1.0 mL, plastic

0354

1

Pipet, 0.5 mL

0353

1

Funnel, plastic

0459

*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.

þ NOTE: For greater accuracy, use laboratory grade pipets.
Equipment needed but not supplied:

1

COD Adapter

5-0087

1

COD Reactor, 12 vial, 120V

5-0102

Or 1

COD Reactor, 12 vial, 23V

5-0102-EX2

Or 1

COD Reactor, 25 vial, 115V/230V

5-0094

Optional Equipment:

1

Volumetric pipet, 2.0 mL

2-2168

2

Volumetric pipet, 5.0 mL

2-2174

1

Volumetric pipet, 0.5 mL

30503

1

Pipet Bulb

2-2164

1

Wipes

2-2069

1

Test Tube Holder

2-2190

Phosphorus in natural waters and wastewaters occurs almost exclusively in the
form of orthophosphates, condensed phosphates (pyro-, meta- and other

Smart2 TEST PROCEDURES 2.04

Phosphorus, Total–HR 1/3