CADMIUM

PAN METHOD • CODE 4017

QUANTITY

CONTENTS

CODE

60 mL

*Buffered Ammonia Reagent

*4020-H

15 mL

Sodium Citrate, 10%

6253-E

30 mL

*PAN Indicator

*4021-G

30 mL

Stabilizing Reagent

4022-G

1

Pipet, 1.0 mL, plastic

0354

2

Pipet, 0.5 mL, plastic

0369

*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.

Cadmium is used in batteries, paint pigments, electroplating processes, and
with other metals in the preparation of alloys. The solubility of cadmium in
natural water is proportional to the hardness or alkalinity of the water.
Cadmium is not an essential nutrient for plants and animals. It is extremely
toxic and can accumulate in the kidneys and liver.

APPLICATION:

Drinking and surface waters; domestic and industrial
wastewater.

RANGE:

0.00–1.00 Cadmium

METHOD:

PAN (1-

[2-Pyridylazo]-2-Naphthol) forms a red complex

with Cadmium (Cd

+2

) at a pH of 10.

SAMPLE
HANDLING &
PRESERVATION:

Analyze sample as soon as possible. If sample must be stored,
acidify with nitric acid to a pH below 2.

INTERFERENCES:

Ag

+2

, Co

+2

, Cu

+2

, Mn

+2

, Ni

+2

, Zn

+2

, Y

+3

, In

+3

Smart2 TEST PROCEDURES 2.04