beautypg.com

LaMotte SMART 2 User Manual

Page 134

background image

PROCEDURE

1.

Press and hold

ON burton until colorimeter turns on.

2.

Press

ENTER to start.

3.

Press

ENTER to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 28 Color) from
TESTING MENU

.

5.

Scroll to and select 28 Color from menu.

6.

Rinse a tube (0290) with color-free water (distilled or deionized water).
Fill to 10 mL line with color-free water.

7.

Insert the tube into chamber, close lid and select SCAN BLANK.

8.

Remove tube from colorimeter. Empty tube.

9.

Rinse tube with sample water. Fill to 10 mL line with water sample.

10.

Insert tube with sample water, close lid and select SCAN SAMPLE. Record
result in color units.

11.

Press

OFF button to turn the colorimeter off or press EXIT button to exit

to a previous menu or make another menu selection.

Color 2/2

Smart2 TEST PROCEDURES 2.04