SELECTING THE VIAL AND WAVELENGTH

The Smart2 Colorimeter has three different vials (the 25 mm 0290 tube,
UDVs and COD tubes) and 4 different wavelengths (430, 520, 570, and
620 nm). The colorimeter uses different settings for each of the twelve
combinations of vial and wavelength. These twelve settings are called
channels. Choose the channel with the correct wavelength and vial for the
test.

Use the

t button to scroll to Select

Vial/WL

and press

Q/ENTER button to select.

EDIT USER TEST01

* Name The Test

Select Vial/WL

New Calibration

Decimal Places

Select Units

Use the

t and s buttons to scroll to the

appropriate channel and press

Q/ENTER button

to select.

NOTE: This is a looping menu.

:

:

:

Ch11 620nm COD

Ch12 570nm COD

SELECT CHANNEL

* Ch1 520nm 25mm

Ch2 430nm 25mm

Ch3 620nm 25mm

Ch4 570nm 25mm

Ch5 520nm UDV

Ch6 430nm UDV

:

:

:

The meter will display the message “Storing”
and return to the EDIT USER TEST01 menu.

Storing

EDIT USER TEST01

* Select The Vial/WL

New Calibration

Decimal Places

Select Units

42