PROCEDURE

1.

Press and hold

ON

button until colorimeter turns on.

2.

Press

ENTER

to start.

3.

Press

ENTER

to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 79 Phosphate H)
from TESTING MENU.

5.

Scroll to and select 79 Phosphate H from menu.

6.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

7.

Insert tube into chamber, close lid and select SCAN BLANK.

8.

Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 2.0
mL of *VM Phosphate Reagent (4410). Cap and mix. Wait 5 minutes for
full color development.

9.

After 5 minute waiting period, mix, insert tube into chamber, close lid and
select SCAN SAMPLE. Record result.

10.

Press

OFF button to turn colorimeter off or press EXIT button to exit to a

previous menu or make another menu selection.