LaMotte SMART 2 User Manual

Page 43

background image

n

ENTERING A NEW CALIBRATION

To enter a new calibration two reacted standards solutions of known
concentration are required: a “low standard” and a “high standard”. These
should be ready to use.

Use the

t button to scroll to New

Calibration

and press

Q/ENTER button to

select.

EDIT USER TEST01

* Select Vial/WL

New Calibration

Decimal Places

Select Units

Input the concentration of the LOW STANDARD
by using the

t and s buttons to scroll the first

digit of the concentration into the first position
on the display. Press

Q/ENTER button to select

that digit (1 for this example).

LOW STANDARD

0

______

* Continue

The number “0” is always the starting point for
the next digit. Continue selecting digits or a
decimal point to enter the concentration (up to
seven characters).

LOW STANDARD

1

0

_____

* Continue

“1.5” has been entered in this example. Press

Q/ENTER button four times to input “0” as the

last four digits. Pressing

Q/ENTER after selecting

the last digit saves the concentration.

LOW STANDARD

1.5

0

___

* Continue

Input the concentration of the HIGH STANDARD
by using the same method as for the low standard.

HIGH STANDARD

0

______

* Continue

SMART2 COLORIMETER 1.07