LaMotte SMART 2 User Manual
Page 43
n
ENTERING A NEW CALIBRATION
To enter a new calibration two reacted standards solutions of known
concentration are required: a “low standard” and a “high standard”. These
should be ready to use.
Use the
t button to scroll to New
Calibration
and press
Q/ENTER button to
select.
EDIT USER TEST01
* Select Vial/WL
New Calibration
Decimal Places
Select Units
Input the concentration of the LOW STANDARD
by using the
t and s buttons to scroll the first
digit of the concentration into the first position
on the display. Press
Q/ENTER button to select
that digit (1 for this example).
LOW STANDARD
0
______
* Continue
The number “0” is always the starting point for
the next digit. Continue selecting digits or a
decimal point to enter the concentration (up to
seven characters).
LOW STANDARD
1
0
_____
* Continue
“1.5” has been entered in this example. Press
Q/ENTER button four times to input “0” as the
last four digits. Pressing
Q/ENTER after selecting
the last digit saves the concentration.
LOW STANDARD
1.5
0
___
* Continue
Input the concentration of the HIGH STANDARD
by using the same method as for the low standard.
HIGH STANDARD
0
______
* Continue
SMART2 COLORIMETER 1.07