LaMotte SMART 2 User Manual
Page 248
PROCEDURE
1.
Press and hold
ON
button until colorimeter turns on.
2.
Press
ENTER
to start.
3.
Press
ENTER
to select TESTING MENU.
4.
Select ALL TESTS (or another sequence containing 96 Tannin) from
TESTING MENU
.
5.
Scroll to and select 96 Tannin from menu.
6.
Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
7.
Insert tube into chamber, close lid and select SCAN BLANK.
8.
Remove tube from colorimeter. Use the plain pipet (0352) to add 4 drops
of *Tannin Reagent #1 (7833). Cap and mix.
9.
Use the 1.0 mL pipet (0354) to add 2.0 mL of *Tannin Reagent #2
(7834). Cap and mix. Wait 30 minutes for full color development.
10.
At end of 30 minute waiting period, mix, insert tube into chamber, close
lid and select SCAN SAMPLE. Record result.
11.
Press
OFF button to turn colorimeter off or press EXIT button to exit to a
previous menu or make another menu selection.
þ NOTES: For best possible results, a reagent blank should be determined to
account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents is obtained.
For the most accurate results, the sample and reagents should be at 20 ±2°C.
Tannin 2/2
Smart2 TEST PROCEDURES 2.04