beautypg.com

LaMotte SMART 2 User Manual

Page 227

background image

PROCEDURE

Use COD adapter.

1.

Preheat COD reactor to 150 ±2°C. Follow safety precautions.

2.

Remove cap from a *Total Phosphorus Acid Reagent Tube (4035). Use a
1.0 mL pipet (0354) to add 5.0 mL of sample.

3.

Use the 0.15 g spoon (0727) and a funnel (0459) to add one level measure
of *Digestion Reagent Powder (4036). Tap funnel to dispense powder
completely. Cap tube tightly and shake until powder completely dissolves.

4.

Place the tube in the COD reactor for 30 minutes.

5.

At the end of the heating period, turn the reactor off. Carefully remove
the tube from the reactor and allow it to cool to room temperature.

6.

At the end of the cooling period, press and hold

ON button until

colorimeter turns on.

7.

Press

ENTER to start.

8.

Press

ENTER to select TESTING MENU.

9.

Select ALL TESTS (or another sequence containing 82 Phos T LR)
from TESTING MENU.

10.

Scroll to and select 82 Phos T LR from the menu.

11.

Carefully remove the cap from the digested tube. Use another 1 mL pipet
(0354) to add 1.0 mL of *Total Phosphorus LR Hydroxide Reagent (4038)
to the tube. Cap and invert to mix.

12.

Wipe the tube with a damp towel to remove fingerprints and smudges.
Wipe with a dry towel.

13.

Insert the tube into the chamber. Select SCAN BLANK. Remove the tube
from the colorimeter.

14.

Use another 1 mL pipet (0354) to add *1.0 mL of Phosphate Acid
Reagent (V-6282). Cap and invert tube to mix.

15.

Use the 0.1g spoon (0699) and a funnel (0459) to add one level spoon of
Phosphate Reducing Reagent (V-6283). Tap funnel to dispense powder
completely. Cap tube and shake until powder dissolves.

16.

Wait 5 minutes.

17.

Wipe the vials with a damp towel to remove fingerprints and smudges.
Wipe with a dry towel.

18.

Insert the tube into the chamber. Select SCAN SAMPLE. Record the result
as Total Phosphorus in mg/L PO

4

.

19.

Press

OFF button to turn the colorimeter off or press EXIT button to exit

to a previous menu or make another menu selection.

þ NOTES: For greater accuracy, use laboratory grade pipets.

Smart2 TEST PROCEDURES 11.04

Phosphorus, Total–LR 3/3