PROCEDURE

1.

Press and hold

ON button until colorimeter turns on.

2.

Press

ENTER to start.

3.

Press

ENTER to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing
94 Surfactants

) from TESTING MENU.

5.

Scroll to and select 94 Surfactants from menu.

6.

Rinse a tube (0290) with sample water. Fill to 10 mL line with sample.

7.

Insert the tube into chamber, close lid and select SCAN BLANK.

8.

Remove the tube from colorimeter.

9.

Use the 0.5 g spoon (0698) to add 0.5 g pH Adjustment Powder (4509).
Cap and mix until powder dissolves.

10.

Use the 0.1 g spoon (0699) to add two measures of Sodium Chloride
Reagent (4877). Cap and mix until powder disintegrates.

11.

Use the 1.0 mL pipet (0354) to add 2.0 mL of *DS Indicator (4508).

12.

Cap and shake for 1 minute.

13.

Wait 5 minutes. DO NOT MIX.

14.

Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result in ppm LAS.

15.

Press

OFF button to turn the colorimeter off or press EXIT button to exit

to a previous menu or make another menu selection.

þ NOTE: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents is obtained.

Surfactants 2/2

Smart2 TEST PROCEDURES 2.04