beautypg.com

LaMotte SMART 2 User Manual

Page 78

background image

PROCEDURE

1.

This test requires a Reagent Blank. Rinse a tube (0290) with clear,
colorless, boron free water. Fill to 10 mL line with clear, colorless, boron
free water.

2.

Use the 1.0 mL pipet (0354) to add 2 mL of *Boron Buffer (4869). Cap
and mix.

3.

Use the 0.15 g spoon (0727) to add one level measure of *Boron Indicator
Powder (4870). Press full spoon against side of jar to compress powder.
Scrape off excess powder on inside neck of bottle. Tap excess off spoon
handle.

4.

Cap and shake vigorously for 30 seconds.

5.

Insert the tube into meter chamber. Close lid.

6.

Start a timer set for 30 minutes. Do not open the lid during the waiting
time. The reaction is photosensitive.

7.

Rinse a clean tube (0290) with Sample Water. Fill to the 10 mL line with
sample water. Repeat steps 2–4.

8.

Insert the tube into the Dark Storage Chamber (0108). Close top.

9.

Start a second timer set for 30 minutes. Do not open the chamber during
the waiting time. The reaction is photosensitive.

10.

When 2 minutes remain on the first timer (Reagent Blank), press and
hold

ON button until colorimeter turns on.

11.

Press

ENTER to start.

12.

Press

ENTER to select Testing Menu.

13.

Select ALL TESTS (or another sequence containing 8 Boron) from
TESTING MENU

.

14.

Scroll to and select 8 Boron from menu.

15.

At the end of the Reagent Blank 30 minute waiting period, remove
Reagent Blank tube from meter chamber. Invert several times to mix.

16.

Insert the tube into meter chamber, close lid and select SCAN BLANK.

17.

Remove the tube from colorimeter.

18.

At the end of the Sample Water 30 minute waiting period, remove
Sample Water tube from Dark Storage Chamber. Invert several times to
mix.

19.

Insert tube into meter chamber, close lid and select SCAN SAMPLE.
Record result in ppm boron.

20.

Press

OFF button to turn the colorimeter off or press EXIT button to exit

to a previous menu or make another menu selection.

Boron 2/2

Smart2 TEST PROCEDURES 2.04