Metrohm IC Net 2.2 User Manual
Page 95
![background image](/manuals/586876/95/background.png)
6.7 709 IC Pump
IC Net 2.2
87
Pressure
Live display of current pressure in MPa.
This field is also available for the
WATCH
WINDOW
.
Flow
Live display of current flow in mL/min. This
field is also available for the
WATCH WIN-
DOW
.
New flow
A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range:
0.01 ... 5.00 mL/min
Min/Max pressure
The
minimum and maximum pressure limit
for the 709 IC Pump can be set in steps of
0.1 MPa.
Entry range:
0.0 ... 50.0 MPa
.
The
set
maximum limit value should lie be-
tween 5 to 10 MPa above the particular
operating pressure or the maximum admis-
sible operating pressure of the column
used. If the pump exceeds the preset limit
value during operation, it is switched off
immediately. At the same time, the
Pressure
field color in the 709 settings window and in
the
WATCH WINDOW
is changed to the
value set for
Out of range
.
The
set
minimum limit value should lie far
enough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolu-
tions of the pump cam due to leaks or inter-
rupted inflow of the eluent, the pump is