6.7 709 IC Pump

IC Net 2.2

87

Pressure

Live display of current pressure in MPa.
This field is also available for the

WATCH

WINDOW

.

Flow

Live display of current flow in mL/min. This
field is also available for the

WATCH WIN-

DOW

.

New flow

A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range:

0.01 ... 5.00 mL/min

Min/Max pressure

The

minimum and maximum pressure limit

for the 709 IC Pump can be set in steps of
0.1 MPa.
Entry range:

0.0 ... 50.0 MPa

.

The

set

maximum limit value should lie be-

tween 5 to 10 MPa above the particular
operating pressure or the maximum admis-
sible operating pressure of the column
used. If the pump exceeds the preset limit
value during operation, it is switched off
immediately. At the same time, the

Pressure

field color in the 709 settings window and in
the

WATCH WINDOW

is changed to the

value set for

Out of range

.

The

set

minimum limit value should lie far

enough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolu-
tions of the pump cam due to leaks or inter-
rupted inflow of the eluent, the pump is