7 Methods
IC Net 2.2
144
To reset the
Run
number press [Ctrl] [F8].
The
Analysis number
window is opened
where the run number for the next run can
be modified to the desired value.
Batch
X/Y: total number of started sample queues
/ current analysis number in the sample
queue (read-only).
7.2.2 Sample
Tab
Sample
of the passport with sample information.
Info 1
First sample description (max. 256 charac-
ters).
Info 2
Second sample description (max. 256
characters).
Volume
Injected volume in
µ
L.
Dilution
Dilution of the sample.
Vial number
Autosampler vial position to take sample
from.
Amount
Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
C
s
= C
c
•
Amount
s
/
Amount
c