10 753 suppressor module, 1 753 suppressor module features, 2 753 suppressor module icon – Metrohm IC Net 2.2 User Manual
Page 105: 3 753 suppressor module window
![background image](/manuals/586876/105/background.png)
6.10 753 Suppressor Module
IC Net 2.2
97
6.10
753 Suppressor Module
6.10.1
753 Suppressor Module features
The Metrohm 753 Suppressor Module consists of a control unit
with a 2-channel peristaltic pump and a block containing a sup-
pressor module.
The 753 Suppressor Module, which is normally operated manu-
ally using the keypad of the instrument or automatically via 732 IC
Detector via remote interface, can be fully operated using the «IC
Net» program. The 753 Suppressor Module is normally con-
nected to two of the event line connections of the 762 IC Interface
using two optional 6.2128.180 cables. Examples for interconnec-
tions with the 753 Suppressor Module can be found in the 762
Instructions for Use.
In order to operate the 753 Suppressor Module with «IC Net»,
the external control must be switched on using the [ Remote ]
key.
For detailed information about the 753 Suppressor Module see
753 Instructions for Use.
6.10.2
753 Suppressor Module icon
The
753 icon
is available in the
SYSTEM
window if a
753 Suppressor
Module
has been installed with the
New system wizard
or by using
the
Setup/New devices/Install new device
option of the
SYSTEM
win-
dow (see 6.1.2).
If the system is connected and the
753 icon
is clicked with the
right mouse button, the following menu appears:
Open
Open
the
753 Suppressor Module
window for parameter
settings (this window can also be opened by double-
clicking the icon).
Unlink
Delete
the
753 icon
from the system.
6.10.3
753 Suppressor Module window
753 icon / Open
The
753 Suppressor Module
window for parameter settings is
opened by selecting this menu option with the right mouse button
or by double-clicking the
753 icon
in the
SYSTEM
window. It con-
sists of the three tabs
Manual
,
Program
, and
Links
.