3 788 ic filtration sample processor window – Metrohm IC Net 2.2 User Manual
Page 138
6 Devices
IC Net 2.2
130
If the system is connected and the
788 icon
is clicked with the
right mouse button, the following menu appears:
Open
Open
the
788 IC Filtration Sample Processor
window for
parameter settings (this window can also be opened
by double-clicking the icon).
Timeout
Open a window for setting the timeout time (in s) for
the scan command.
Unlink
Delete
the
788 icon
from the system.
6.18.3
788 IC Filtration Sample Processor window
788 icon / Open
The
788 IC Filtration Sample Processor
window for parameter set-
tings and manual control is opened by selecting this menu option
with the right mouse button or by double-clicking the
788 icon
in
the
SYSTEM
window. It consists of the four tabs
Manual
,
Program
,
Configuration
and
Links
.
Manual
The manual control of the 788 IC Filtration Sample Processor can
be done on the
Manual
subtab of the
Manual
tab which is only
available for a connected system.
Move
Start moving the rack to place the specified
vessel at the needle.