8 update calibration – Metrohm IC Net 2.2 User Manual
Page 179
![background image](/manuals/586876/179/background.png)
7.5 Calibration and quantification
IC Net 2.2
171
Base
Base for calculations (
Area
or
Height
) in
quantification and calibration procedures.
Indicates, which parameter is to be used as
a peak response.
Reference channel
Channel which is used to measure area or
height (read-only).
Formula
Formula for calibration function. There are
six possible calibration dependencies for
linear and nonlinear calibration curves.
Statistical weight
Weighting parameter used to weight cali-
bration points deviations on calculating co-
efficients of the calibration formula. Apart
from the uniform weighting
1
the values
1/x
,
1/x
2
,
1/y
and
1/y
2
are also possible.
Standard component
Name of standard component in the case
of
Internal standard calibration
or
Tabulated
calibration
method (details see on-line help).
Concentration of standard
Concentration of standard component in
the case of
Internal standard calibration
or
Tabulated calibration
method (details see on-
line help).
7.5.8 Update
calibration
IC NET / Method / Calibration / Update
Selection of this menu item opens the
Update
window.
Retention time
Replace the expected retention times of the
components with the retention times of the cor-
responding peaks obtained in the current run.
Indexes
Fill
Index
field for components with zero index.
Coefficients
Recalculate all calibration coefficients for all
components (performs recalibration).
Coef. (balance)
Recalculate coefficient of the universal com-
ponent. The aim of this action is to adjust the re-
sponse factor of this component so, that the total