Metrohm IC Net 2.2 User Manual
Page 188
7 Methods
IC Net 2.2
180
As long as no universal component is de-
fined, only peaks with non-zero concentra-
tions are listed.
The peak table includes the following col-
umns:
Peak number
,
Retention time
,
Height
,
Area
,
Response factor
,
Concentration
,
Con-
centration%
,
Name
Relative concentration
This quantification method uses the internal
standard method to calculate the relative
concentrations of the components. For this
procedure, a
Standard component
must be
selected and the concentration of this inter-
nal standard must be entered in the
Con-
centration of std
field.
As long as no universal component is de-
fined, only peaks with non-zero concentra-
tions are listed.
The peak table includes the following col-
umns:
Peak number
,
Retention time
,
Height
,
Area
,
Response factor
,
Relative concentration
,
Relative concentration%
,
Name
Index
This column reports retention indexes of
identified components.
The peak table includes the following col-
umns:
Peak number
,
Retention time
,
Width/2
,
Height
,
Index
,
Name
Column test
Column test quantification method calcu-
lates multiple values that are necessary to
evaluate column performance.
The peak table includes the following col-
umns:
Peak number
,
Retention time
,
K'
(ca-
pacity factor),
TP
(number of theoretical
plates per column),
TP/m
(number of theo-
retical plates per m),
HETP/dp
(reduced
height equivalent to theoretical plate),
Asym.
(peak asymmetry),
Name
Custom
This quantification method enables to cus-
tomize the peak table on the user's choice
using the
button.
The peak table for the methods supplied
contains the following columns as stan-
dard:
Peak number
,
Retention time
,
Height
,
Area
,
Conc.
,
Name
Standard component
Internal standard component for quantifica-
tion procedure.