Metrohm IC Net 2.2 User Manual
Page 54
![background image](/manuals/586876/54/background.png)
5 Interfaces
IC Net 2.2
46
Range
Range for AD converter.
Example: converts
±
2500 mV to 2
24
bits
Entry range:
2500, 1250, 625, 312.5, 156.25,
78.125, 39.062 mV
Read current parameters from 771 Com-
pact Interface.
Send current parameters to 771 Compact
Interface.
RS settings
The
RS settings
tab of the
771 Compact Interface
window contains
RS232 parameters for the 771 Compact Interface.
Baudrate
Baud rate of the device.
Selection:
1200...115200
Default value:
57600
Parity
Parity check.
Selection:
None
,
Even
,
Odd
Default value:
None
Handshake
Enable/disable software handshake mode.
Selection:
Swchar
,
none
Default value:
SWchar