6.18 788 IC Filtration Sample Processor

IC Net 2.2

133

Num

Row number (read-only).

Task

Program instruction (see below).

Parameter

Parameter for program instruction (see be-
low).

Enabled

Enable program start (a disabled program
is not started).

Add new program instruction.

Delete selected program instruction.

List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

tab:

Instruction Parameter entry

Meaning

Move sample,

sample+1,

spec.#, 1...n

Move rack to place the specified

vessel at the needle.

Lift

shift, work, rinse,
special,
0, 1...125 mm

Raise or lower the lift to the speci-

fied position.

Pump

on, off, 1...999 s

Switch on/off peristaltic pump at

the tower.

Scan "scan

command"

Scan remote interface until the

specified signal is received. In ad-

dition to the normal scan com-

mands (details see 788 Instruc-

tion for Use) special commands

defined on the

Configuration

tab

can be used.