6.4 761 Compact IC
IC Net 2.2
69
6.4.4 Hardware
settings
Selection of the
Hardware
menu item of the
761 icon
using the right
mouse button opens the hardware settings window consisting of
the three tabs
Hardware
,
Outputs
and
Alarm stops
.
Hardware
The
Hardware
tab of the hardware settings window defines gen-
eral parameters which are set automatically at power on of the
761 Compact IC.
Remote lines after power on
The remote output lines 1
…
8 are set to this
values after power on or a manual stop with
Shutdown hardware
.
Selection:
0
,
1
Thermostat
Operating temperature of the conductivity
cell.
Selection:
25
,
30
,
35
,
40
,
45 °C
,
off
Thermostatting functions only if the ambient tempera-
ture is at least 5 °C lower than the operating tempera-
ture. It normally takes 30
…
60 min after power on until
a temperature stability of
±
0.01 °C is attained.
Flow correction
Factor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range:
0.9
...
1.09
The flow correction is determined by meas-
urement of the actual flow rate with the aid
of a measuring cylinder as follows:
Flow correction
=
Displayed flow rate
Measured flow rate