Metrohm IC Net 2.2 User Manual
Page 72
![background image](/manuals/586876/72/background.png)
6 Devices
IC Net 2.2
64
even without connection to the PC.
Entry range:
0.0
...
25.0
MPa
Full scale,
µ
S/cm
Startup value for full scale range.
Selection:
50
,
250
,
1000
µ
S/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1
Initial
The
Initial
tab of the
761 Compact IC
window for settings of system
startup values is only available for a disconnected system.
Flow, mL/min
Startup value for flow rate of the high-
pressure pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit
for high-pressure pump.
Entry range:
0.0
...
25.0 MPa
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump.
Entry range:
0.0
...
25.0
MPa
Full scale,
µ
S/cm
Startup value for full scale range.
Selection:
50
,
250
,
1000
µ
S/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1
Time program
On the
Program
tab of the
761 Compact IC
window a user-defined
time program for the remote output lines of the 761 Compact IC
can be entered. This program is started automatically as defined
in the
Start mode
window (see section 4.4.3) either at the moment
the determination is started (
Start with determination
) or at the
moment the sample is injected (
Start with inject
).