Metrohm IC Net 2.2 User Manual
Page 84
![background image](/manuals/586876/84/background.png)
6 Devices
IC Net 2.2
76
Oven
Switching on/off column oven and setting
operating temperature.
Range:
15 ... 60 °C
Range
Selection of measuring range:
pA
10 ... 5000 pA
.
nA
0.1 ... 50 nA
.
µA
0.01 ... 5 µA
.
Full
scale
The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
depend on the preset measuring
Range
.
Filter
Electronic damping of the analog output
signal.
Selection:
0.1, 0.2, 0.5, 1, 2, 5 s
Offset
Offset of the zero point for the analog out-
put signal.
Selection:
-50 ... +50 %
(in steps of 10 %)
Actual
Cell potential
Display of current cell potential in V
Oven
Display of current oven state.
Range
Display of current measuring range.
Full
scale
Display of current full scale range.
Filter
Display of current damping constant.
Offset
Display of current offset.
Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (
*.smt
) as long as the file is
not saved.
Pulse mode
The
Pulse mode
subtab of the
Method parameters
tab on the
817
Bioscan
window is used to set the parameters for the pulse mode.
This window is only available if
Mode
has been set to
Pulse
on the
Manual
tab.
Method data
Cell
Definition of three different working poten-
tials, which are applied to the amperometric
detector cell cyclically. The set potentials
are graphically displayed if the
Method
op-
tion is enabled above the graphic window.
E1, E2, E3
Working potentials.
Range:
-2 ... +2 V
.
t1, t2, t3
Time duration to apply the working poten-
tial.
Range:
0.01 ... 2 s