6.6
733 IC Separation Center
6.6.1
733 IC Separation Center features
The Metrohm 733 IC Separation Center is a thermally and elec-
tronically isolated wet part which accommodates injectors, col-
umns, detectors, suppressor module and pulsation dampener
and is controlled by and connected to the 732 IC Detector. The
following versions are available:
•
2.733.0010 IC Separation Center with 1 injector
•
2.733.0120 IC Separation Center with 2 injectors, metal-free
•
2.733.0130 IC Separation Center with 1 injector and 1 sup-
pressor module, metal-free
The 733 IC Separation Center, which is normally operated by the
732 IC Detector, can be fully operated using the «IC Net» pro-
gram.
The 733 IC Separation Center with one injection valve is nor-
mally connected to the 732 IC Detector using the 6.2125.090 ca-
ble. For the operation of a 733 IC Separation Center with two in-
dependent columns and two injection valves, two individual
732 IC Detectors are needed. Each of this detectors is connected
to the 733 IC Separation Center using a 6.2125.090 cable. For
operation of a two-valve system with «IC Net», two
732 icons
and
two
733 icons
(one for valve A and one for valve B) have to be in-
stalled in the
SYSTEM
window. To install the 733 IC Separation
Center in a system the
733 IC Separation Center
device must be se-
lected, the type of Separation Center is detected automatically.
If the Separation Center is connected to the 762 IC Interface,
one 6.2115.070 cable for every event line is needed. The number
depends on the type of Separation Center, i.e. for the 733.0130 IC
Separation Center three event lines are necessary to control the
valve (fill, inject) and the suppressor (step). In this case the ap-
propriate
733.0XX0 Separation Center
device must be selected to
install the 733 IC Separation Center in a system.
For detailed information about the 733 IC Separation Center see
732/733 Instructions for Use. Examples for interconnections
with the 733 IC Separation Center, the 732 IC Detector and the
762 IC Interface can be found in the 762 Instructions for Use.