Metrohm IC Net 2.2 User Manual

Page 173

background image

7.5 Calibration and quantification

IC Net 2.2

165

min C (max C)

Minimum (Maximum) concentration
value for the component. Components
whose concentration leaves the range

min C

max C

are marked in the peak table

by the sign "

!

".


Additionally, the component table contains the following buttons:

Add a new component (an empty column)
to the component table.

Delete the current component from the
component table.

Open

the

Peak identification

window (see

section 7.5.5).

Open

the

Concentration table

(see section

7.5.6).

Open the component window with display
of calibration curve and calibration parame-
ters.

Accept all changes and close the compo-
nent table.

Reject all changes and close the compo-
nent table.

The standard methods supplied with the «IC Net» program all
contain a component table with pre-defined components with en-
tries for the names and retention times.

If the output of all peaks is required in the result report then a
so-called universal component can be defined in the compo-
nent table for this purpose; this receives the value

0

in the

Time

field and no peak is allocated to it. The name entered under

Name

(e.g.

unknown

) is then used for all peaks for which no

component can be allocated.