4 762 ic interface window – Metrohm IC Net 2.2 User Manual
Page 47
![background image](/manuals/586876/47/background.png)
5.2 762 IC Interface
IC Net 2.2
39
5.2.4
762 IC Interface window
762 icon / Open
The
762 IC Interface
window for parameter settings is opened by
selecting this menu option with the right mouse button or by click-
ing the
762 icon
on the toolbar. It consists of three tabs
Init
,
RS set-
tings
, and
Links
.
Init
The
Init
tab of the
762 IC Interface
window contains data acquisi-
tion parameters for the 762 IC Interface.
Label
Optional label to name the interface with
maximum 8 characters.
ROM version
Instrument program version number.
Measuring per second
Number of data points measured per s.
Entry range:
10, 20, 30, 50, 60 points/s
Range
Range for AD converter.
Example: converts
±
2500 mV to 2
24
bits
Entry range:
2500, 1250, 625, 312.5, 156.25,
78.125, 39.062 mV
Second press on RUN/STOP means 'stop' when measuring
If this option is enabled, the running data
acquisition will be stopped if the RUN/STOP
button on the 762 IC Interface front panel is
pressed. This is equal to
Stop determination
.
If this option is disabled, the running data
acquisition will not be stopped if the
RUN/STOP button on the 762 IC Interface