Metrohm IC Net 2.2 User Manual
Page 51
5.2 762 IC Interface
IC Net 2.2
43
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction Parameter
entry Meaning
Events
0, 1, p, *
Set event output lines 1...7 to the
desired values. For entry of the first
value, enter
1
,
0
,
p
or
*
. For entry of
the other values, move the cursor in
front of the value to be changed
and enter
1
,
0
,
p
or
*
.
Program END,
RESET
The
END
flag can be used to end a
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The
RESET
flag is used to reset the
parameters to the system startup
values.
Events configuration
On the
Events configuration
subtab user-defined event commands
can be defined, which can be inserted into a time program after
being activated with
.
Name (1
st
column)
User-definable name of the event com-
mand (e.g.
Start_766
).